Steps to move the magnet current from 74.51 A to 74 A

  1. Select the option "Up to 74.51 A at 0.5 A/m" from the "Presets" field.


  2. Press the "To Set" button and wait until the "Output current (A)" increases up to 74.51 A. It will take about 5 mins, and once the output current reaches 74.51 A, it will be automatically switched back to the "Hold".


  3. Press the toggle button next to the "Main Switch Heater" status indicator. Once you press it, then it will show a similar screen as shown below.
    Image Added

    Once this process is completed, then the "PD Readings" will indicate ON (see the screenshot below).
    Image Added

  4. Select the option "Down to 74 A at 0.5 A/m" from the Presets.
    Image Added

  5. Press "To Set".
    Image Added

    Wait until the magnet current moves to 74 A, and it will go back to "Hold".
    Image Added

  6. Now you can take the NMR baseline. Proceed with step #2 on How to take a baseline and then take data on UVA-NMR system



Steps to move the magnet current from 74 A to 74.51 A after taking the baseline


  1. Select the option "Up to 74.51 A at 0.5 A/m" from the "Presets" field.
    Image Added

  2. Press "To Set", and wait until the current goes up to 74.51 A and also it automatically switch to "Hold".
    Image Added
  3. Press the toggle button next to the "Main Switch Heater" status indicator. Once you press it, then it will show a similar screen as shown below.
    Image Added

    Once it is completed, then "PD Reading" indicators will indicate "OFF".
    Image Added
  4. Press "Hold" button and wait for a couple of seconds.
  5. Press "To Zero", and wait until the Shim current goes to Zero.
    Image Added
  6. It will go back to the "Hold" once the shim current is ramped-down.
    Image Added
  7. Now, you can continue NMR data taking.SS