- Ensure the "AUTO" control of the GV-Bypass PID is ON if you are in the Hold Mode.
- Click "Full Open" the GV-Bypass (wait until you can see the limit switch for OPEN is GREEN). "Auto Control with Pressure" will be automatically switched to OFF.
- Wait until the pressure is stabilized (here you have to coordinate with the Target Expert to find which pressure you will need it to be stabilized).
- Open GV (Gate Valve), and wait until the fridge pressure is stabilized.
- Fully close the GV-Bypass valve.
- Turn ON the boosters of the roots-stack (from bottom to top ). Ensure to wait 1 min per each pump when you turn them ON one after the other.
- Crack open the RUN valve by 0.54 turns (look for Fridge Valves VI on the CCP).
- Turn ON the microwave following Microwave Operation.