  1. Ensure the "AUTO" control of the GV-Bypass PID is ON if you are in the Hold Mode.
  2. Click "Full Open" the GV-Bypass (wait until you can see the limit switch for OPEN is GREEN). "Auto Control with Pressure" will be automatically switched to OFF.
  3. Wait until the pressure is stabilized (here you have to coordinate with the Target Expert to find which pressure you will need it to be stabilized).
  4. Open GV (Gate Valve), and wait until the fridge pressure is stabilized.


  5. Fully close the GV-Bypass valve.
  6. Turn ON the boosters of the roots-stack (from bottom to top ). Ensure to wait 1 min per each pump when you turn them ON one after the other.
  7. Crack open the RUN valve by 0.54 turns (look for Fridge Valves VI on the CCP).
  8. Turn ON the microwave following Microwave Operation.