Checklist for Unloading the insert without NH3/ND3

Item #LocationCheck Item
1Control room areaDid you read the procedure?
2Control room areaHave you notified the Target Expert, the run coordinator, and the FNAL RCT team? Is the FNAL RCT team available to access the cave with you?
Do you have NM3 and NM4 keys?
3Control room areaAre the Gate Valve (GV), and the GV-Bypass valve closed?
4Control room areaIs the target lifter in the bottom position?
5Control room areaPPE:
Do you have cryo-gloves?
Are you ready with PPE level 3 (Tyvex suit, gloves, shoe covers)?
6Control room areaIs there a Target Operator available at the counting-house during this process? Make sure that the TO has a walkie-talkie to communicate.
7Control room areaHave you had a job briefing with everyone in the access team?
8Material handling stationAre the cart, collar+screws, insert-handle, and two nuts for the insert-handle ready?
9Material handling stationDo you have the wrench to disconnect the microwave? and also a 3/16 wrench in case NMR connectors are tight?
There should be a plastic bag (which contains those tools) hanging on the cart.
10On the way to NM3

Have you turned ON the fans?
Have you OPENED the backfill gHe supply line (+ the valve D) on the cryoplatform?
Have you turned ON the lights in the cave?

11In-cave (before unloading)Have you opened the two hand valves on the backfill panel?
12In-cave (before unloading)Do you have a banc-off flange, and an extra seal ready?
13In-cave (before unloading)Have you installed the insert handle, and connected the hook to the insert's eyebolt?
14In-cave (before unloading)Have you disconnected the microwave wave-guide and NMR cable(s) and arranged them properly?
15In-cave (before unloading)Have you checked the 0.5 psig relief valve's venting?
16In-cave (after unloading)Have you properly closed the fridge using a banc-off flange and a seal?
17In-cave (after unloading)Have you closed the two hand valves on the backfill panel?
Have you confirmed whether the blanc-off and its seal closed the fridge properly?
18On the way back from NM3

Have you turned OFF the lights in the cave?
Have you turned OFF the fan?
Have you CLOSED the backfill gHe supply line (+ the valve D) on the cryoplatform?



Checklist for Loading the insert without NH3/ND3

Item #LocationCheck Item
1Control room areaDid you read the procedure?
2Control room areaHave you notified the Target Expert, the run coordinator, and the FNAL RCT team? Is the FNAL RCT team available to access the cave with you?
Do you have NM3 and NM4 keys?
3Control room areaAre the Gate Valve (GV), and the GV-Bypass valve closed?
4Control room areaIs the fridge level below 20%? If not, bring it down below 20% using the annealing heat
5Control room areaIs the target lifter in the bottom position?
6Control room areaPPE:
Do you have cryo-gloves?
Are you ready with PPE level 3 (Tyvex suit, gloves, shoe covers)?
7Control room areaIs there a Target Operator available at the counting-house during this process? Make sure that the TO has a walkie-talkie to communicate.
8Control room areaHave you had a job briefing with everyone in the access team?
9Material handling stationAre the insert, cart, collar+screws, insert-handle, and two nuts for the insert-handle ready?
10Material handling stationDo you have the wrench to disconnect the microwave? and also a 3/16 wrench in case NMR connectors are tight?
There should be a plastic bag (which contains those tools) hanging on the cart.
11On the way to NM3

Have you turned ON the fans?
Have you OPENED the backfill gHe supply line (+ the valve D) on the cryoplatform?
Have you turned ON the lights in the cave?

12On the way to NM3Have you installed the insert handle, and connected the hook to the insert's eyebolt?
13In-cave (before loading)Have you opened the two hand valves on the backfill panel?
14In-cave (before loading)Do you have an extra seal ready?
15In-cave (before loading)Have you checked the 0.5 psig relief valve's venting?
16In-cave (before loading)Have you properly opened the fridge banc-off flange (and a seal)?
17In-cave (after loading)Have you closed the two hand valves on the backfill panel?
Have you confirmed whether the insert-top with the seal closed the fridge properly?
18On the way back from NM3

Have you turned OFF the lights in the cave?
Have you turned OFF the fan?
Have you CLOSED the backfill gHe supply line (+ the valve D) on the cryoplatform?