You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 4 Next »

  1. Ensure the "AUTO" control of the GV-Bypass PID is OFF.
  2. Fully open the GV-Bypass (wait until you can see the limit switch for OPEN is GREEN).
  3. Wait until the pressure is stabilized (here you have to coordinate with the Target Expert to find which pressure you will need it to be stabilized).
  4. Open GV (Gate Valve), and wait until the fridge pressure is stabilized.
  5. Fully close the GV-Bypass valve.
  6. Turn ON the boosters of the roots-stack.
  7. Crack open the RUN valve by 0.54 turns.
  8. Turn ON the microwave following Microwave Operation.
  • No labels