There are two paths for the fridge return flow.
1. Gate Valve Bypass (GV-bypass)
Options:
1. Absolute moves by percent: Enter the set point (in %) and press "ENTER". The boxes to the right allow you to do increments in terms of +/-1, +/-5, +/-10.
2. Absolute moves by turns : Enter the number of turns set-point, and press "ENTER". The boxes to the right allow you to do increments in terms of +/-0.01, +/-0.05, +/-0.1.
3. Auto Control with Pressure: Click the "Toggle Switch" shown in the Purple box (see below) to enable the PID control on the Fridge Pressure by providing the set-point in Torr.
4. Fully Open/Close : Use the "Full Open" button or "Full Close" button as needed (see the two yellow arrows towards the upper-left of the following picture).
2. Gate Valve (GV)
IMPORTANT: Please check with the Target Expert on Shift before opening the GV.
* Do not open the GV,
(1) if the fridge is at atmospheric pressure or above (0.5 psi)
and
(2) if the GV-Bypass was not opened before.
* If the fridge is above atmospheric pressure (or ~0.5psi), then turn OFF the roots-pumps while the Rotary Vane pump (backing pump) is still running. After that, while keeping the GV closed, open GV-Bypass valve with the steps of 20% until the fridge pressure is below 100 Torr, and then fully open the GV-Bypass.
Configurations
Configuration | "Power" switch | "Valve" switch | Indications |
---|---|---|---|
Close | Down | Down | "Valve Close" → Green "Valve Open" → OFF |
Open | Up | Up | "Valve Close" → OFF |
Configuration when the GV is "Closed"
Configuration when the GV is "Open"