See Teams of SpinQuest Target Group page for the contact information


Contact Persons

SystemOn-call personsQTIshara, Kenichi, Liliet, FarooqNMRSyed, Farooq, Ishara, Liliet, Forhad, VaniyaMagnet System (in-cave piping)Farooq, Erensto, VibodhaSlow ControlsKenichi, Huma, VibodhaTarget LifterVibodhaMicrowaveVibodhaTarget Material RelatedFarooqInventoryFarooq
SystemProblemPossible CauseSteps to do
QTIf the QT-HMI got frozen
How to fix the QT-HMI if its screen got frozen

If QT-HMI has an alram window opened and showing the codes 603 and 602It's a known warning from some software development → close the windowHR3 Tripped Warning Message

If the Dewar level indicators have large fluctuationsLevel sensor tubes are contaminatedFixing dewar level indicator
FNAL Approved procedure: https://seaquest-docdb.fnal.gov/cgi-bin/sso/ShowDocument?docid=10401

Coldheads blocking alarmPossible contamination
or cold heads have been running so long without any intake gHe for a while
  1. Resolving Liquefier Blocking Alarm when Liquefier is on Liquefy Mode
  2. Resolving Liquefier Blocking Alarm when Liquefier is on Idle Mode
IVCIf the Turbo pump stopped running
IVC (← check this page)