Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

MicrowavesOFF
NMRUsually, it should be on "Take Data" (check with the Target Expert on Shift if there are issues)
Fridge
  • Only the Backing Pump (Rotary Vane pump) should be running, and the rest of the roots pumps should be OFF
    See Roots pumps PLC interface page
  • Maintain the fridge LHe level between 50% and Heat-Ex Bottom. In other words, let the fridge level drops down to 50% and then refill it up to Heat-Ex Bottom by opening the run-valve (then close it).
  • The fridge pressure should be on PID control at 600 400 Torr (see the table below).
  • Adjust the fridge RUN/Bypass valve between 0.5 turns (for a slow fill) or 2 turns (for a relatively quick fill), and close it when the fridge fill is complete and continue to monitor the fridge level and refill when the level reaches back to 50%.
    See these links
    (a) How to Open or Close Fridge Valves
    (b) How to identify the fridge LHe level is at Heat-Ex Bottom
  • Discuss with the Target Expert when to switch from the Hold Mode to the Production Mode: Transitioning from the Hold mode to the Production Mode
Magnet
  • Ensure the magnet has enough LHe.
  • Ensure Magnet LN2 level is above 50%, if not refill it.
QT System
  • Ensure the purifier LN2 level and the magnet shield LN2 levels are above 50%, if not refill it.
    See Helium Liquefier (QT) System Summary page to identify where to look at for those LN2 levels.
  • Ensure both liquefiers produce LHe with the rate of ~ 4 L/ hr (with an uncertainty +/- 0.3). If not, then contact the Target Expert on shift. You can use https://e906-gat1.fnal.gov/data-summary/e1039/target-par-preset/target-LHe-prod-rate.php
  • Check LCW outlet Temperatures from the cold-head compressors and ensure all those temperatures are below  95 F. If any of those are. > 100 F then, report to the Target Expert immediately and follow Target Expert's instructions.

...



How to check and control
Fridge

Check with the Target Expert among the following options:
(1) Set the gv-bypass PID controlled for maintaining 570 400 Torr by Auto Control with Pressure
(2) Fully open gv-bypass

How to control the Fridge Return flow (GV and GV Bypass)
Separator20 slmHow to control the Separator Return flow

...