In this mode, you will need to coordinate with the Target Expert on shift regarding the planned duration of time spent in this mode.
This mode will be alternated with the Production Mode during normal operations with microwaves to keep up with the LHe production from the QT system.

Configuration / Things to maintain

MicrowavesOFF.
NMRTaking DataUsually, it should be on "Take Data" (check with the Target Expert on Shift if there are issues)
Fridge
  • Only the Backing Pump (Rotary Vane pump) should be running, and the rest of the roots pumps should be OFF
    See Roots pumps PLC interface page
  • Maintain the fridge LHe level between 50% and Heat-Ex Bottom. In other words, let the fridge level drops down to 50% and then refill it up to Heat-Ex Bottom by opening the run-valve (then close it).
  • The fridge pressure should be on PID control at 400 Torr (see the table below).
  • Adjust the fridge RUN/Bypass valve
(to ~0
  • between 0.5 turns (for a slow fill) or 2 turns (for a relatively quick fill)
to keep
  • , and close it when the fridge fill is complete and continue to monitor the fridge
LHe
  • level
around 50%, and then close the run valve.
MagnetEnsure the magnet has enough LHe.
Magnet
  • Ensure the magnet has enough LHe.
  • Ensure
implement Switching from Production Mode to Hold Mode
Ensure
  • Magnet LN2 level is above 50%, if not refill it.
QT System
  • Ensure the purifier LN2 level
is
  • and the magnet shield LN2 levels are above 50%, if not refill it.
    See Helium Liquefier (QT) System Summary page to identify where to look at for those LN2 levels.
  • Ensure both liquefiers
are producing

Outlet Flows

...

  • Check LCW outlet Temperatures from the cold-head compressors and ensure all those temperatures are below  95 F. If any of those are. > 100 F then, report to the Target Expert immediately and follow Target Expert's instructions.


Outlet Flows



How to check and control
Fridge       8

Check with the Target Expert among the following options:
(1) Set the gv-bypass PID controlled for maintaining 400 Torr by Auto Control with Pressure
(2) Fully open gv-bypass

How to control the Fridge Return flow (GV and GV Bypass)
Separator       20 slmHow to control the Separator Return flow
Magnet     ~7 (on pressure PID at 0.9 psi)(link to a page)


Other useful instructions:

Steps for Transitioning from the Hold mode to the Production Mode




################# I think the links below are too much information for the Target Operator manual, can be suitable for the target expert manual ####

Conditions

  1. Fridge parameters when the fridge is in Hold Mode
  2. QT system conditions when the fridge is in Hold Mode
  3. Magnet conditions when the fridge is in Hold Mode
  4. Cryo-control parameters / conditions when the fridge is in Hold Mode
  5. Microwave parameters / conditions when the fridge is in Hold Mode
  6. Target Lifter conditions when the fridge is in Hold ModeUVA-NMR conditions when the fridge is in Hold Mode