  1. Check the magnet current. If it is not 74.51A, then go to step #2. If the current is 74.51A, then change it to 74A.

  2. On NMR Terminal, click "NMR Tune".
    Image Added
  3. Click the box next to the label "Baseline"; and a new window will be opened.
    Image Added
  4. Click "Create a new Baseline", then this new window will be automatically closed.
    Image Added
  5. Change the number of sweeps to 5000.
    Image Added
  6. Then Click "One Point" button, it will take about ~320 seconds for a 5000 sweeps baseline.
    Image Added
  7. Wait until the indicator (RED) goes to "Pause" after taking the baseline.
    Image Added
  8. Click "New Baseline" button (to select the baseline that you just took), then it will open a new window).

  9. Select the Baseline from drop-down list on the left (use the time-stamp to select the correct one), and press the "Selected Highlighted Baseline"; then the window will be closed automatically.
    Image Added
  10. Change the number of sweeps to 100.
  11. Click "Take Data".
    Image Added



If you want to create a new "event" file (for example: to take TE measurement), follow the steps on How to create a new event file on PDP























The following section explains the NMR terminal, polarization, and NMR Signal and how to take a new baseline and data. 

...

Insert 5000 at point 1 (Sweeps) and click at point 2 on 'One Point.' The baseline will be completed in around 324 seconds. You will see a new baseline at the bottom left of the above picture.


How to Take TE Measurement: 

TE Calibration with the SpinQuest Polarized Target System - Wiki Post Portal - Confluence (virginia.edu)