This page is dedicated to the Roots pump operating manual. The Root Pump has three Boosters (WH7000-1, WH7000-2, and WH7000-3) and one backing pump (SV630BF). To turn each Booster on or off, we have to take a 40-second timelapse.
Please follow the steps to turn ON and OFF Root pump boosters.
Turning ON Root Pump Boosters:
To turn on root pump boosters, we must follow the sequence from Booster 3 → Booster 2 → Booster 1 and wait 40 seconds for each Booster.
- Turn ON Booster 3 by pressing the OFF button; it will turn ON Booster 3.
- Please wait for 40 seconds, and you see their rotational speed becomes 70 Hz.
- Turn ON Booster 2 by pressing the OFF button; it will turn ON Booster 2.
- Please wait for 40 seconds, and you see their rotational speed becomes 70 Hz.
- Turn ON Booster 1 by pressing the OFF button; it will turn ON Booster 1.
- Please wait for 40 seconds, and you see their rotational speed becomes 70 Hz.
Please see the following picture after turning on all three Root Pump Boosters.
Turning OFF Root Pump Boosters:
To turn off root pump boosters, we must follow the sequence from Booster 1 → Booster 2 → Booster 3 and wait 40 seconds for each Booster.
- Turn OFF Booster 1 by pressing the ON button; it will turn OFF Booster 1.
- Please wait for 40 seconds, and you see their rotational speed becomes 0 Hz.
- Turn OFF Booster 2 by pressing the ON button; it will turn OFF Booster 2.
- Please wait for 40 seconds, and you see their rotational speed becomes 0 Hz.
- Turn OFF Booster 3 by pressing the ON button; it will turn OFF Booster 3.
- Please wait for 40 seconds, and you see their rotational speed becomes 0 Hz.
Please see the following picture after turning off all three Root Pump Boosters.