We sometimes need to (nearly) empty the fridge in a short time, for example;
- To load/unload the target insert, and
- To consume LHe in the fridge and/or the magnet fast.
Conditions
- You have been trained by an experienced person on this procedure.
- The nose LHe level is above 80%. If it is below 80%, this method cannot evaporate LHe effectively, and thus you should simply wait until LHe evaporates by itself.
- The heat tape for the turret is in operation.
- About the roots pump, the rotary vane pump (SV630BF) is ON and the boosters (WH7000-1, 2 & 3) are OFF.
- The gate-valve bypass is fully opened (at 0.25 turns).
- On Fridge_Valve_Main.vi;
- The mode of the level probe is "CONT" (not "HOLD").
- The Interlock with Heat-Ex Top is enabled.
- The "Voltage" knob of the Annealing Plate Control Box is fully rotated clockwise.
Procedure
- Turn ON the "Power Control" switch of the Annealing Plate Control Box, and turn it OFF when the top-section pressure of the roots pump reaches 10 Torr.
- Never leave the switch ON longer than required in this and later steps.
- You had better keep your finger on the switch, not to forget turning it OFF.
- Confirm that the main return flow increases to 100-150 SLM. If it is higher/lower, find any difference in the conditions.
- Turn ON the "Power Control" switch, and turn it OFF when the top-section pressure reaches 16 Torr.
- Confirm that the main return flow increases to 150-250 SLM. If it is higher/lower, find any difference in the conditions.
- Wait until the top-section pressure decreases to 8 Torr (or below).
- Repeat steps 3-5 until the nose LHe level becomes ~80%.
- If you need consume LHe in the magnet, fill the fridge from the magnet.
Change the separator flow to 25 SLM.
- If the separator flow was 7 SLM, you might wait for 0.5-1.5 hours. To be optimized.
- Open the fridge bypass valve at 3.0 turns. Confirm that the fridge return flow goes up to ~100 SLM.
- Monitor the LHe level probe and the chip resistors to confirm that LHe is being transferred normally.
- Close the fridge bypass valve (i.e. set it at 0 turns) when the Heat-Ex Top resistor goes up to ~1400 Ohm.
- Change the separator flow to 20 SLM.
Repeat steps 1-7 until the fridge and/or magnet levels reaches what you need.
- Change the separator flow to 7 SLM.
- Change the opening of the gate-valve bypass to 0.050 turns.