Location and Function

The fridge (or refrigerator) is inserted into the inner space of the target magnet from the top.  It takes LHe from the magnet LHe space thru its "separator" and stores LHe in its bottom-most space called "nose".  The nose space is cooled down to 1 K by the helium evaporation method.  The evaporated helium is pulled by the roots pump thru the gate valve or the gate-valve bypass.

Sensors

Operations


Important Facts

  1. How long does the fridge would last if the starting liquid level is 100%? ~ 8 hrs from (100% to 10%) with maintaining cernox 7 temperature below 50 K
    Note: Set Separator flow to 10slm.

    https://e906-gat1.fnal.gov/data-summary/e1039/target-all.php?sel_mode=date_time&Y0=2024&M0=06&D0=16&h0=20&m0=00&s0=00&Y1=2024&M1=06&D1=17&h1=11&m1=10&s1=00&Cryo+Temperature_Cernox+7&Cryo+Level_Nose+LHe+Level


  2. How long does the fridge would last if the starting liquid level is Heat Ex. Bottom?  ~16 hrs
  3. How long would the fridge be cold if the level probe reaches zero?  ~1.5-2.0 hrs


Procedures

  1. Fridge Fill Procedure
  2. Fridge Emptying Procedure