If the magnet is already in the "persistent mode", then the VI should look like the following picture.
Steps to move the magnet current from 74.51 A to 74 A
- Select the option "Up to 74.51 A at 0.5 A/m" from the "Presets" field.
- Press the "To Set" button and wait until the "Output current (A)" increases up to 74.51 A. It will take about 5 mins, and once the output current reaches 74.51 A, it will be automatically switched back to the "Hold".
- Press the toggle button next to the "Main Switch Heater" status indicator. Once you press it, then it will show a similar screen as shown below.
Once this process is completed, then the "PD Readings" will indicate ON (see the screenshot below). - Select the option "Down to 74 A at 0.5 A/m" from the Presets.
- Press "To Set".
Wait until the magnet current moves to 74 A, and it will go back to "Hold". - Now you can take the NMR baseline. Proceed with the step #2 on How to take a baseline and then take data on UVA-NMR system