  1.  Switch the LIquefier to Manual + Liquefy mode (Manual + Idle mode is also the same because Liquefy/Idle option has no effect when it is in the 'Manual' mode)
  2.  Set the inlet flow to 60slm: enter 60slm on FC501(A/B)Q (then press enter)
  3.  Set the outlet flow for the maximum flow by opening SVV(A/B) and the manual hand valve bypassing the SVN(A/B)
     Note: Currently, this bypass valve exists only for LQ-B (will install one for LQ-A soon)
  4.  Turn ON the cold heads manually on "Helium Compressors" screen