- Ensure the "AUTO" control of the GV-Bypass PID is OFF.
- Check the Present Position on the VI. If it is already closer to 100% and also whether the limit switch for OPEN is GREEN. If not, then fully open the GV-Bypass (wait until you can see the limit switch for OPEN is GREEN).
The following is an example screenshot when the GV-Bypass is fully opened
![Image](https://media.discordapp.net/attachments/1192955920976711762/1246780176294215800/image.png?ex=665da206&is=665c5086&hm=d3077234a46236d60892510b53870c0f71c42a5582a4c200a1776158b0652dec&=&format=webp&quality=lossless&width=1520&height=968)
- Wait until the pressure is stabilized (here you have to coordinate with the Target Expert to find which pressure you will need it to be stabilized).
- Open GV (Gate Valve), and wait until the fridge pressure is stabilized.
![](/download/attachments/170902013/image-2024-6-2_6-1-39.png?version=1&modificationDate=1717326099147&api=v2)
- Fully close the GV-Bypass valve.
- Turn ON the boosters of the roots-stack (from bottom to top ).
- Crack open the RUN valve by 0.54 turns.
- Turn ON the microwave following Microwave Operation.