Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

The fridge is equipped with two valves (run valve and bypass valve) to control the incoming helium from the magnet.  We control the valves using the Cryo Control Panel (CCP) on the target computer.

Fridge Valve Panel in CCP

  • You can open the Panel for the Fridge Valves (Fridge_Valve_Main.vi) by clicking the "Open" button for "Fridge Valve" (Image Added) on the CCP Main Panel (CCP_Main.vi).
  • Interface for control and monitoring
    • Run valve = Right middle section of the Fridge Valve Panel.
    • Bypass valve = Right bottom section.
  • The number of valve turns is used to specify the opening of valves.
    • Fully closed = 0 turns
    • Fully opened = 3 turns

Image Added


How to Move Run or Bypass Valve

The operation procedure is common to the run and bypass valves. 

  1. Make sure which valve you handle, and make sure you use the correct interface (i.e. the middle or bottom section).
  2. Decide the set point of valve turns that you move to.
  3. Either
    1. Enter the set point into the "Set Turns" text box (Image Added) and hit the "Enter" key, or
    2. Press the "Set Turns" button(s) as many as needed (Image Added).

In Case of Mistake and Unexpected Situation

  • If you enter a wrong set point, you might press the "Stop Move" button (Image Added) to stop the valve motion.
  • If you set the valve turn out of the valid range (0-3 turns) somehow, please contact the target expert immediately.  In principle the set point is limited to the valid range by software (VI).