Navigation:
Press F5 > Display Settings (Press F5 again, If you want to return to the main screen)
(After pressing F5: you can navigate among the settings pages, and adjust values)
F1 / F2 > Navigate to the Right / Left screens
F3 / F4 > Increase / Decrease the parameter on the selected screen
Press F5 once you are done, to get back to the regular (green) display.
Operational Settings
Parameter | Value |
---|---|
Buffer Tank Max Pressure | 250 psig |
Buffer Tank Min Pressure | 20 psig |
SV7 Opening Pressure Threshold | 90 psig |
SV7 Closing Pressure Threshold | 70 psig |
Max rise in pressure to stop recovery cycle | 1 psig |
SV7 Idle behavior (SV7 stays closed when compressor "Idle") | YES |
Sample time to test pressure rise | 15 mins |
Times between samples | 15 mins |
On Ishara changed both the Sample time period (minutes), and Sample Spacing (minutes) to 15 minutes.
...