Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

Configuration / Things to maintain

MicrowavesOFF.
NMRTaking Data
FridgeOnly the Backing Pump (Rotary Vane pump) should be running, and the rest of the roots pumps should be OFF
Adjust fridge RUN/Bypass valve (to ~0.5 turns) to keep the fridge LHe level around 50%, and then close the run valve.
MagnetEnsure the magnet has enough LHe.
Discuss with the Target Expert when to implement Switching from Production Mode to Hold Mode
Ensure Magnet LN2 level is above 50%, if not refill it.
QT SystemEnsure the purifier LN2 level is above 50%, if not refill it.
Ensure both liquefiers are producing LHe with the rate of ~ 4 L/ hr.




Outlet Flows


Outlet Flow (in slm)How to check and control
Fridge       8How to control the Fridge Return flow
Separator       20How to control the Separator Return flow
Magnet     ~7 (on pressure PID at 0.9 psi)(link to a page)

...