In this mode, the fridge level should be maintained above the 'Heat Exchanger Bottom' (see the Purple curve on the top-left graph of the screenshot below), and the separator is kept at cold with 20slm return flow. IMPORTANT: If you see the Heat-Ex Top resistance (purple curve) starts increasing, then reduce the number of turns on the fridge run-valve using the "Fridge Valve Panel" VI on the CCP


Configuration / Things to maintain

Sub-systemStatus
MicrowavesON
NMRTaking Data
FridgeAll roots pumps are running
Adjust fridge RUN/Bypass valve to keep the fridge LHe level
MagnetEnsure the magnet has enough LHe.
Discuss with the Target Expert when to implement Switching from Production Mode to Hold Mode
Ensure Magnet LN2 level is above 50%, if not refill it.
QT SystemEnsure the purifier LN2 level is above 50%, if not refill it.



Outlet FlowsOutlet Flows (nominal)


Outlet Flow (in slm)Run Valve / Bypass ValveHow to check and control
Fridge       Nominal 45slm (can be higher
(depends on the microwave power + beam)
Fridge Run valve → 1 turn
Bypass valve → CLOSED
How to control the Fridge Return flow
Separator       20 N/AHow to control the Separator Return flow
Magnet     ~7 (on pressure PID at 0.9 psi)N/A(link to a page)

...