Keep monitoring the target parameters, say Insert#7, Insert #8, Annealing A, and Annealing B. Where Insert #7 and Insert# 8 are the Cernox (CX-1030) sensors on the Top and Bottom cup of the target insert. Where Annealing A and Annealing B are the metal temperature in the fridge, monitor these four target parameters. Where Insert#7 and Insert# 8 are in resistance (Ohm) and Annealing A and Annealing B are in temperature (K). Please see a screenshot below for these target parameters: 


Below is the conversion table for the Cernox (CX-1030) from temperature resistance. 


We have two target inserts: Insert # 1 and Insert # 2. Each insert has two Cernoxs attached to the top and bottom cups, and each Cernox is calibrated. Please check with Target Expert which Insert is loaded in the fridge because each Cernox has a different temperature and resistance curve. Please see below the temperature and resistance curves for each Insert:

Insert # 1: 

1. Insert # 7 (Target Parameter) Cernox Serial Number X130217 installed on the Bottom CupPlease see the respective temperature resistance curves with their serial number below.  

1. Insert # 8 (Target Parameter) Cernox Serial Number X130192 installed on the Bottom Cup. Please see the respective temperature resistance curves with their serial number below.  



Insert # 2: 

1. Insert # 7 (Target Parameter) Cernox Serial Number X133422 installed on the Bottom CupPlease see the respective temperature resistance curves with their serial number below.  

...