Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

1. Insert # 7 (Target Parameter) Cernox Serial Number X130217 installed on the Bottom Cup. Please see the respective temperature resistance curves with their serial number below.  

1. Insert # 8 (Target Parameter) Cernox Serial Number X130192 installed on the Bottom Cup. Please see the respective temperature resistance curves with their serial number below.  



Insert # 2: 

1. Insert # 7 (Target Parameter) Cernox Serial Number X133422 installed on the Bottom CupPlease see the respective temperature resistance curves with their serial number below.  

1. Insert # 8 (Target Parameter) Cernox Serial Number X133421 installed on the Bottom Cup. Please see the respective temperature resistance curves with their serial number below.  

Image AddedImage Added








If you see the target parameters, Insert #7 and Insert #8 start lowering down to around 132 Ohms, which is around 50K, then open the Run valve at 0.55 or 0.65 turns to get the liquid in the fridge nose. Keep monitoring the resistance of sensors; it should start increasing after opening the Run valve. Wait for 30 minutes. If the temperature exceeds ~80K (lower than 101 Ohm), open the run valve at two turns, set the separator flow at 30slm, and call the target expert. 

...