Below is the conversion table for the Cernox (CX-1030) from temperature resistance. 


We have two target inserts: Insert # 1 and Insert # 2. Each have


If you see the target parameters, Insert #7 and Insert #8 start lowering down to around 132 Ohms, which is around 50K, then open the Run valve at 0.55 or 0.65 turns to get the liquid in the fridge nose. Keep monitoring the resistance of sensors; it should start increasing after opening the Run valve. Wait for 30 minutes. If the temperature exceeds ~80K (lower than 101 Ohm), open the run valve at two turns, set the separator flow at 30slm, and call the target expert. 

...