QT system

FrequencyTask
Daily
  1. Check HR3 oil level
  2. Check P40 and P41 pressure gauges (if the pressure difference is > ~60 psi, then you will have to arrange to clean the purifier)
  3. Check the LCW flow into the cold-head compressors for the past 12 hrs. If there is an unusual behavior please notify about it on the discord.
  4. Check LCW temperatures from the cold-head compressor outlets for the past 12 hrs. If there is an unusual behavior please notify about it on the discord.
Weekly
  1. F1, F2 filters on HR3. See →  Cleaning the filters F1, F2 at HR3 for the steps
  2. Check MV101 filter on top of the HR3. For cleaning, follow → Cleaning the filter at MV101



    






Target Expert/Helper/Trainee needs to update this spreadsheet: https://docs.google.com/spreadsheets/d/1k836pi3-3F1O7qrVbh-RQXJcwsCeYpV61wKufTBvMig

...