- Turn off HR3 (Kaeser) Compressor.
- Close MV107 (you will need to get the key from the key box located at the west-wall of the control-room to open the lock).
- Close MV101
- Crack open MV103 to vent with a slow flow
Note: Turn ON the fans at NM4-Hall to avoid an ODH alarm. - Once all gHe is vented, then open the part of the filter and clean it with kim-wipes and alcohol, then let it dry before putting it together.
- Re-install the filter's cup and make sure it is firm.
- Take the Walkie-Talkie and once person should go to the outside tank's supply tank.
- Crack open MV107 (outside), and communicate to the person in the hall.
- The person in the hall can rotate the bottom part of that filter and bleed a few times, and close it.
- Gently open MV107 until is fully opened.
- Bleed gHe from MV103 a few times and close MV103.
- Open MV101.
{"serverDuration": 53, "requestCorrelationId": "7e3501fcf1a95c13"}