This measurement tells us how well the purification of gHe from the purifier. Therefore, that value should in principle be closed to the purity of a cylinder (see: Reference gHe purity sample from a gHe cylinder )
- Ensure V30, V35 are open (see the image below).
- Ensure V9a and V21a are open.
- Ensure SV03 is open.
- Navigate to either Liquefier A or Liqiefier B to ensure PT501 reads a pressure higher than 10 psig.
- Open "SVSAMPLE" on the QT HMI.
- Open "REGPM" and then open "NVPM" to have some gHe flow.
- Connect the portable purity meter's probe to this port on the QT HMI cabinet.
- Once the purity measurement is taken, then close the valves in the following order.
NVPM > REGPM > SVSAMPLE - Measure the purity, and take a picture of the reading, send it to the target-expert (or to the target-group), make a elog entry (with date, time, location, and purity).