- Open the gHe cylinder located at the west wall (right-most cylinder)cylinder, and regulate ~10 psig.
- Open "Vfill".
- Open and close Vvent multiple times to flush, then close Vvent.
- Open "SVREF" on the QT HMI screen.
- Open "REGPM" and then open "NVPM" to have some gHe flow.
- Connect the portable purity meter's probe to this port on the QT HMI cabinet.
- Once the purity measurement is taken, then close the valves in the following order.
NVPM > REGPM > SVREF - Open "Vvent" to vent the pressure in the purge-line from the cylinder while keeping the "Vfill" open.
Ensure the pressure indicator "on-the-left" on that regulator on the cylinder shows "0" (zero). - Close "Vvent".
- Close "Vfill"
- Close the cylinder.