IMG_0254.jpg

  1. HR3 Manual
  2. HMI Interface and Connectivity Manual
  3. Liquefiers (A B) Manual
  4. Purifier Manual
  5. QT System P&ID Information
  6. Coldhead Compressors Manual
  7. HMI Cabinet Manual
  8. Pneumatic Air Supply Manual

Check this page before and after the system is restarted → Operational Configuration

Routine Maintenance Procedures


General/Other Procedures

PurposeInstructions/Links
System Startup Procedure

QT System Starting Procedure

QT PLC (only) restart procedure

How to restart the QT PLC

Cooldown

Look At Updated Page

For planned power outage

Shut Down and Restart Procedures

Taking purity measurements at various places in the system

Measuring gHe purity


Troubleshoots

IssueReasonWhat to do?
If QT-HMI has an alram window opened and showing the codes 603 and 602It's a known warning from some software development → close the windowHR3 Tripped Warning Message
If the QT-HMI got frozen
How to fix the QT-HMI if its screen got frozen
If the Dewar level indicators have large fluctuationsLevel sensor tubes are contaminatedFixing dewar level indicator
FNAL Approved procedure: https://seaquest-docdb.fnal.gov/cgi-bin/sso/ShowDocument?docid=10401
Coldheads blocking alarmPossible contamination
or cold heads have been running so long without any intake gHe for a while
  1. Resolving Liquefier Blocking Alarm when Liquefier is on Liquefy Mode
  2. Resolving Liquefier Blocking Alarm when Liquefier is on Idle Mode
If FCA and FCB shows strange values (like 71.11slm on both)Probably a consequence of radiation damageHow to fix if the outlet flows of Liquefiers are showing strange values
If the coldhead compressors show Alarm due to high temperatures of LCWHigher outside temperaturesHow to clear the coldhead compressor LCW temperature alarm

Useful Information

  1. The volume of an outside gHe tank: 1600 cubic feet (V = 45306.95 L)
  2. 100 slm rate on Magnet FC is equivalent to ~ 192 LHe liters per day (i.e. assuming both LQ A and B are running)

Useful Contact Information


  1. Cryofab : https://www.cryofab.com/contact
    Call +1 (800) 426-2186 and ask for extension #14 (Mr. Tam)
    Tell this serial number and he will recognize us
    Image
  • No labels