Purpose | Instructions/Links |
---|---|
Cleaning (Liquefier) | |
Cleaning (Purifier) | |
Cleaning (HR3 piping) | Purging HR3 after a maintenance |
Cleaning the buffer tank | Cleaning the Buffer Tank on the Cryoplatform |
Cleaning the filters F1, F2, F3, and the filter before MV101 | Cleaning the filters F1, F2 at HR3 Cleaning the filter at MV101 |
Fills | Filling LN2 to the purifier using the outside LN2 tank |
Taking purity measurements at various places in the system |
Purpose | Instructions/Links |
---|---|
System Startup Procedure | |
QT PLC (only) restart procedure | |
Cooldown | |
For planned power outage | |
Taking purity measurements at various places in the system |
Issue | Reason | What to do? |
---|---|---|
If QT-HMI has an alram window opened and showing the codes 603 and 602 | It's a known warning from some software development → close the window | HR3 Tripped Warning Message |
If the QT-HMI got frozen | How to fix the QT-HMI if its screen got frozen | |
If the Dewar level indicators have large fluctuations | Level sensor tubes are contaminated | Fixing dewar level indicator FNAL Approved procedure: https://seaquest-docdb.fnal.gov/cgi-bin/sso/ShowDocument?docid=10401 |
Coldheads blocking alarm | Possible contamination or cold heads have been running so long without any intake gHe for a while | |
If FCA and FCB shows strange values (like 71.11slm on both) | Probably a consequence of radiation damage | How to fix if the outlet flows of Liquefiers are showing strange values |
If the coldhead compressors show Alarm due to high temperatures of LCW | Higher outside temperatures | How to clear the coldhead compressor LCW temperature alarm |