- While you are seeing the error message, press the left-most button to clear the error message on the screen (see "1" label).
- Once the error message is gone, then press the right-most button to start the pump (see "2" label). Once the pump is started, you can notice that its rotation speed start increasing up to 1000 Hz.
- If the error message stays after performing the step #1, then turn OFF the pump using its power switch on its side bottom, wait for 10 seconds and turn the same switch ON, and perform the step #2.
Error codes
If the pump behaves strangely and if you are not able to restart the pump, perform a Hard-ResetÂ
How to do a Hard-Reset on the HiCube pump
- Turn off the main power switch on the side of the pump.
- Unplug the main power connection from the pump.
- Wait until the pump controller screen completely turns off. (Even without power connected, the pump screen will be active for about 5 minutes.)
- Connect the power connection.
- Turn on the main switch.
- Wait until the controller boots up (Sometimes you will see a non-zero value indicating the current rotation speed)
- Press the rightmost button on the controller to start the pump (see "2" label in the first picture).
- Make sure the rotation speed starts increasing up to 1000 Hz.