The set point for the auto control is 600 Torr. Note: if you are setting this at the end of a production run, and want to take data for polarization decay constant (T1), then use 23 Torr instead. Check with the Target Expert regarding this, otherwise use 600 Torr.
Ensure, the auto control on the pressure PID of the GV-Bypass VI is enabled.
10. Set the separator flow to 10slm.
Procedure for Filling the Fridge with LHe
The fridge is supposed to have no LHe at the beginning. You can/should use this preset page to monitor the key parameters.
A serious failure during this procedure is an overfill of LHe to the fridge (above the separator). When you lose track of the liquid level, you must close the bypass valve (on Fridge_Valve_Main.vi).
Change the separator flow to 25 SLM (See → How to control the Separator Return flow), and send a discord message to the group mentioning that you are changing the separator flow and will fill the fridge soon.
Wait until the Ann-Plate Top resistor goes up to 1400 Ohm, which indicates that the liquid started accumulating. It usually takes 30 minutes from step #3. If it takes longer than 40 minutes, search for unexpected conditions.
Wait until the nose LHe level reaches 100%. It usually takes 15 minutes from the previous step.
Wait until the Heat-Ex Bottom resistor goes down from 1400 Ohm and then up. It usually takes 10 minutes from the previous step.
Wait until the Heat-Ex Top resistor goes up to 1400 Ohm, which indicates that the liquid is as high as Heat-Ex Top. It usually takes 10 minutes from the previous step.
Close the bypass valve.
Usually the software interlock sets it at 0.5 turns automatically.
Change the level-probe mode back to "HOLD" if you changed it in step 5.
This attached PDF version of the Fridge filling procedure is exactly the same as we have above, along with the annotated pictures. Filling_Procedure.pdf
The following procedure is the same as the above procedure but written in such a way with annotated pictures.