This page is to collect all information about the Slow Controls system. Old information is saved in this sub page.
Procedure for turning off and on the slow control rack, in case of planned/unplanned power outage.
n/a
Two types of alarm systems for the target group (not the shifter) are available at present.
Subsystem | Variable | Unit | LOLO | LO | HI | HIHI | Enabled? | Note |
---|---|---|---|---|---|---|---|---|
Cryo Control Panel | IVC Pressure | Torr | n/a | n/a | 1e-6 | 1e-5 | No | |
Roots-Pump LCW | W1/W2/W3 | GPM | 0.2 | 0.3 | n/a | n/a | Yes | |
FSW1 | GPM | 1.2 | 1.3 | n/a | n/a | Yes | ||
QT HMI | Liq-A/B LHe level | L | ?? | ?? | 180 | 200 | No | |
PT8 | psi | 30? | 40? | 100? | 110? | No | ||
Purifier LN2 level | L | 40 | 50 | 85 | 95 | No | ||
Last Updated: | 2022-Nov-22 11:06 |
Sensor Device | N | Sensor Location | Controller Device | Controller Location | Front End | Notes |
---|---|---|---|---|---|---|
TH HFC-303 | 2 | Magnet & Separator | THCD-400 #1 | Slow-control rack | CCP @ Target computer | |
TH HFM-305 | 1 | Fridge | THCD-400 #1 | ditto | CCP @ Target computer | |
??? | 5 | roots LCW inflow | Arduino | Magnet rack | Target computer | Photo wanted. |
??? | 1 | NM4 LCW inflow | ACNET | n/a | Target computer?? | Photo |
Note about installation:
Item number | Sensor Device | N | Sensor Location | Controller Device | Controller Location | Front End | Note |
---|---|---|---|---|---|---|---|
9 | Pfeiffer PKR 361 | 1 | Attached to dewar | Pfeiffer TPG 361 | Slow-control rack | Target computer | |
8 | MKS 722B #1 | 1 | Return manifold | MKS 946 #1, ch. 1 | Slow-control rack | Target computer | |
5 | MKS 722B #2 | 1 | Main return line, after roots pump | MKS 946 #1, ch. 2 | ditto | ditto | |
1 | MKS 690 #1 (100 Torr) | 1 | Fridge via flex hose | MKS 670 #1 | Slow-control rack | Target computer | |
2 | MKS 690 #2 (1000 Torr) | 1 | Neat MKS 690 100 Torr | MKS 670 #2 | Slow-control rack | Target computer | |
3 | MKS 615 | 1 | He3 pressure probe | MKS 670 #3 | Slow-control rack | Target computer | |
6 | Pfeiffer TPR280 | 1 | Separator return line, near return manifold | Maxigauge # 1, ch. 3 | Magnet Rack | Target computer | To be replaced with APR250 |
7 | Pfeiffer TPR280 | 1 | Magnet return line, near return manifold | Maxigauge # 1, ch. 4 | Magnet Rack | Target computer | To be replaced with APR250 |
x | Pfeiffer APR250 | 1 | Main return line, near return manifold | Maxigauge # 1, ch. 5 | Magnet Rack | Target computer | To be replaced with another sensor |
10 | Pfeiffer TPR280 | 1 | Roots pump, LANL | Maxigauge # 1, ch. 1 | Magnet Rack | Target computer | |
11 | Pfeiffer TPR280 | 1 | Roots pump, LANL | Maxigauge # 1, ch. 2 | Magnet Rack | Target computer |
Item | Location | Notes |
---|---|---|
9 | Already installed | |
8 | Pipe fitting: KF-40 Sensor fitting: 4-VCR Female Flex hose (needed because the gas port of the sensor must not be directed upward) KF40 to KF16 converter : https://locoscience.com/products/kf40-to-kf16-flanges-vacuum-fitting-conical-reducer-adapter KF16 flex hose: https://locoscience.com/products/kf16-flexible-vacuum-bellow-1000mm?_pos=1&_sid=63838aa66&_ss=r | |
5 | Pipe fitting: KF-16 Sensor fitting: 4-VCR Female Adapter: KF-16 to 4-VCR Male https://www.idealvac.com/Adapter-KF-16-to-4-VCR-Male/pp/P101374 Location: towards upside of the white piping (see the left image) | |
1 & 2 | Pipe fitting: 1000Torr (~ $ 470) KF16 to 1/2" swagelok | |
3 | This is He Vapor pressure sensor (MKS 615) | Waiting for piping |
6 & 7 | ![]() ![]() | 6) Separator line from cave should be connected to: Tee: https://locoscience.com/products/3-way-tee-kf25-flange-vacuum-fitting 7) LHe vapor line from Magnet Dewar should be connected to: Adapter: https://locoscience.com/products/kf25-flange-to-1-npt-male-vacuum-fitting-adapter flex hose: https://www.amazon.com/ISO-KF-Corrugated-Bellow-Aluminum-Centering/dp/B09QGWGM2W (Qty 4) Also, need a count of various converters we already have. |
Fermilab Tech will help us install this one | ||
10 & 11 | LANL sensors. Check with Misha | |
4 (AUX) | ![]() | MKS 722B (cylinder silver) seen in the photo has been moved to the roots pump area for #5. TPR 280 will be place for #4 (AUX). We need a longer cable. |
Sensor Device | N | Sensor Location | Sensor in place | Cable in place | Controller Device | Controller Location | Controller in place | Front End | Notes |
---|---|---|---|---|---|---|---|---|---|
Vishay Chip 1K OHM | 8 | Inside of Fridge | No | Not sure | LakeShore 218 #1 | Slow-control rack | Yes? | Target computer | |
6 Vishay Chip 1 K OHM + 2 cernox | 8 | Target Insert | no | Not sure | LakeShore 218 #2 | Slow-control rack | Yes? | Target computer | |
Type-T TC | 8 | Magnet coils | Yes | Yes | MCC E-TC #1 | Slow-control rack | Yes | Target computer | |
Type-T TC | 3 | LHe QT fill line (2 vacuum layer, 1 on stinger) | Yes | Yes | MCC E-TC #2 | Slow-control rack | Yes | Target computer | |
Type-T TC | 2 | Magnet IVC | Yes | Yes | MCC E-TC #2 | ditto | ditto | ditto | |
Type-T TC | 1 | Separator line | Yes | Yes | MCC E-TC #2 | ditto | ditto | ditto | |
Type-T TC | 1 | Top seal of fridge | Yes | Yes | MCC E-TC #2 | ditto | ditto | ditto | |
Type-T TC | 2 | Microwave cooling (1 on EIO, 1 on cooling water) | Yes | Yes | MCC E-TC #3 | Slow-control rack | Yes | Target computer | |
Type-T TC | 2 | Fridge for Waqar's manual annealing system | Yes | Yes | MCC E-TC #3 | ditto | ditto | ditto | |
Type-T TC | 2 | Fridge for Henery's automated annealing system | No | No | Arduino?? | Annealing Panel | No | ?? | |
Type-K TC | 10 | Surface of 10 cold heads | Yes | Yes | LabJack U3-LV | Slow-control rack | Yes | Target computer | |
List of devices to be included
Controller | Sensor | Interface | Controller+sensor ready? | Standalone VIs ready? | VIs integrated into CCP? | Condition for FNAL Review? |
---|---|---|---|---|---|---|
THCD 400 | 2 x HFC + HFM | RS-232 | Yes | Yes | Yes |
|
AMI 1700 | LHe4 level probe | RS-232 | Yes | n/a | Yes |
|
Run valve | Motor driver: AM ST5-S ADC: MCC USB-202 | RS-232 USB | Yes Yes | n/a | Yes |
|
Bypass valve | Motor driver: AM ST5-S ADC: MCC USB-202 | RS-232 USB | Yes Yes | n/a | Yes |
|
MKS 670 | MKS 690 (100 Torr) | GPIB | Yes+No (*2) | n/a | Yes |
|
MKS 670 | MKS 690 (1000 Torr) | RS-232 | Yes | Yes | Yes |
|
MKS 670 | MKS 615 | GPIB | Yes+No (*3) | n/a | Yes |
|
MKS 946 | 2 x MKS 722 | RS-232 | Yes | Yes | Yes |
|
MaxiGauge | 5 x Pfeiffer Vacuum TPR 280 | RS-232? | Yes | Yes | Yes |
|
This is table contain the status of the slow control rack and magnet rack: RackStatusApril2022.xlsx
Some (Not all) priority:
Date | Title | Presenter | File |
---|---|---|---|
| VIs for MCC E-TC | Kenichi | labview_20210810.pdf |
| Update on LN2 Bos | Vibodha | SCMeetig_August25.pptx |
| Zulkaida | ||
| Step count ADC length relationship | Vibodha | |
| Updates on slow control rack and target computer | Kenichi | labview_20210915.pdf |
| MKS 670 + PDP and LakeShore 218 + GPIB | Kenichi | labview_20210922.pdf |
| VIs for MCC E-TC | Kenichi | labview_20210929.pdf |
| VIs for MCC E-TC | Kenichi | labview_20211013.pdf |
| Lifter Vi update | Vibodha | |
| MCC E-TC & Slow Control Data Monitor | Kenichi | labview_20211103.pdf |
| Slow Control Data Monitor & MCC USB-202 | Kenichi | labview_20211110.pdf |
| Slow Control Data Monitor & MCC E-TC with long TC | Kenichi | labview_20211117.pdf |
| Some list about pressure sensor | Zulkaida | |
| Slow Control Data Monitor, NMR Computer & MCC USB-202 with TC | Kenichi | labview_20211124.pdf |
| NMR computer & MKS 615 | Kenichi | labview_20211201.pdf |
| Lifter VI | Vibodha | |
| Slow-Control System, NMR Computer & MKS 615 | Kenichi | labview_20211208.pdf |
| Test of MKS 615 with LN2 Test of MCC E-TC + Long TC in LN2 | Kenichi Kenichi | |
| TCP/IP communication of spill info, N of sweeps/measurement on PDP | Kenichi | slow_cont_20220119.pdf |
| PDP Readout Test | Kenichi | slow_cont_20220126.pdf |
| TCs for magnet coils and others; Rearrangement of KF connection | Kenichi | slow_cont_20220216.pdf |
| Cryo Control Panel | Kenichi | slow_cont_20220302.pdf |
| TCs for magnet coils; Cryo Control Panel | Kenichi | |
| Cryo Control Panel | Kenichi | |
| Cryo Control Panel | Kenichi | |
| Run+Bypass Valve in CCP | Kenichi | |
| MKS670 and MaxiGauge in CCP | Kenichi | |
| LCW flow meter, TC, CCP | Kenichi | |
| Measurement of NMR noise level Logging, Automated alarm | Kenichi Kenichi | |
| QT HMI | Kenichi | |
| QT HMI, Chart-axis control | Kenichi | |
| TCs; Cables for MKS 615 | Kenichi | |
| Magnet PS Shim Relay | Kenichi | |
| Logging of roots-pump PLC; Pressure-sensor locations | Kenichi |