You are viewing an old version of this page. View the current version.

Compare with Current View Page History

« Previous Version 2 Current »

  1. Press "To Zero", to confirm that CCP can communicate with the magnet PS.


  2. Press the Toggle button of the main switch heater.  It turns on the main switch heater and also the two shim switch heaters.
    1. When you use the PS front panel, you have to press additionally the two buttons of the shim switch heaters on CCP, to turn them on.




  3. Confirm that the two LEDs on the shim PS are ON (if you can go into the hall).
    Image

  4. Set "2 A/m" and "45 A".
    1. You can use the "Presets" selector, or enter the values into the text boxes.
      not re

  5. Press "To Set" and wait until the current becomes the set point (SP) and "HOLD".  It takes 23 minutes.


    After ~23 minutes.


  6. Set "1 A/m" and "63 A".


  7. Press "To Set" and wait until the current becomes the SP and "HOLD".  It takes 18 minutes.


    After ~18 minutes


  8. Set "0.5 A/m" and "74.47 A".


  9. Press "To Set" and wait until the current becomes the SP and "HOLD".  It takes 24 minutes.


    After ~24 minutes



  10. Switch to the persistent mode. Follow the steps in the next section.

Reference: UVA elog 154

  • No labels