Operations
The following procedures are applicable to the remote control interface (i.e. Cryo Control Panel; CCP). When you use the front panel of the magnet PS, some button operations are different.
To Ramp Up
- Confirm that the shim PS is powered ON (if you can go in the hall).
- Press "To Zero".
- Press the Toggle button of the main switch heater. It turns on both the main and shim switch heaters.
- When you use the PS front panel, you press additionally the two buttons of the shim switch heaters on CCP, to turn them on.
- Confirm that the two LEDs on the shim PS are ON (if you can go into the hall).
- Set "2 A/m" and "45 A".
- You can use the "Presets" selector, or enter the values into the text boxes.
- Press "To Set" and wait until the current becomes the set point (SP) and "HOLD". It takes 23 minutes.
- Set "1 A/m" and "63 A".
- Press "To Set" and wait until the current becomes the SP and "HOLD". It takes 18 minutes.
- Set "0.5 A/m" and "74.78 A".
- Press "To Set" and wait until the current becomes the SP and "HOLD". It takes 24 minutes.
- Switch to the persistent mode, if you want.
Reference: UVA elog 154
To Switch to Persistent Mode
- Press the Toggle button of the main switch heaters. It turns off both the main and shim switch heaters.
- Press "To Zero" and wait until the current becomes 0 A and "HOLD". It takes 4 minutes, where the ramp rate is slow at the beginning.
To Switch from Persistent Mode
- Confirm that the shim PS is powered ON. If not, turn it on and check why it was off.
- Confirm that the main and shim switch heaters are OFF. If not, something is wrong.
- Set "74.78 A", using the Presets selector. The ramp rate becomes "0.5 A/m" but it will not matter.
- Press "To Set" and wait until the current becomes the SP and "HOLD". It takes 2.5 minutes.
Confirm that "Clamped Rate Limit Quench" is not red.
To Ramp Down
- Switch back from the persistent mode, if necessary.
- Press the Toggle button of the main switch heater. It turns on both the main and shim switch heaters.
- Set "0.5 A/m" and "63 A".
- Press "To Set" and wait until the current becomes the SP and "HOLD". It takes 24 minutes.
- Set "1 A/m" and "45 A".
- Press "To Set" and wait until the current becomes the SP and "HOLD". It takes 18 minutes.
- Set "2 A/m" and "0 A".
- Press "To Set" and wait until the current becomes the SP and "HOLD". It takes 23 minutes.
- Press the Toggle button of the main switch heater. It turns off both the main and shim switch heaters.
Reference: UVA elog 174
Documents
- ORC 2049 - ORC of Super conducting magnet electronic connectors (in the target cave)
- oxford_seppo_magnet_cryostat_handbook.pdf
- oxford_seppo_work_notes.pdf
t.b.w.