Slow Pre-cooling:
1. Enable the Flow PID control of the Magnet Return Bypass at 65 slm.
2. Open VJVA/B, open VJVT and close VPC.
Note: Liquifier A/B should be on Liquify mode. If at the time of starting the slow pre-cooling the Liquifier A/B is on Idle mode, start the slow pre-cooling and when 4L are consumed on the Liquifier then change it to Liquify mode.
...
1. Open VJVA/B (ensure VPC is already opened).
2. Wait for TX1 temperature to drop bellow below 100K (~ 4 min with GV02 fully opened).
3. Enable the Flow PID control of the Magnet Return Bypass at 65 slm.
4. Open VJVT and wait for magnet pressure to increase to 6 PSI.
5. Close VPC and proceed to the filling steps.
...
1. Open VJVA/B (ensure VPC is already opened).
2. Wait for TX1 temperature to drop bellow below 100K (~ 4 min with GV02 fully opened).
3. Enable the Flow PID control of the Magnet Return Bypass at 65 slm.
4. Open VJVT and close VPC.
Proceed to the filling steps anytime after the transfer lines are cold. The system is sufficiently cold once the Tank T temperature decreases after opening VJVT (about 1 hour).
...