Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

We get the best transfer efficiencies when the transfer lines from the Liquifier to the magnet are sufficiently cold. We have tested two modes (paths) to pre-cool the system: the slow and the fast pre-cooling as well as a combination of these two modes which allows us to reduce the slow pre-cooling time.

Image Modified             

...

Image Modified

The fast pre-cooling transfer efficiencies are about 47-49% which does not allow filling the entire magnet volume using just one Liquifier. Nevertheless, higher efficiencies of about 54% can be obtained with the fast pre-cool if the Liquifier pressure is reduced to 3.5 PSI at the expense of nearly 2 hours (~8L) where the Liquifier is not producing while reaching the set pressure. With the fast pre-cooling the last part of the the transfer line is not re-cooled (see Figure 1). Although the fast pre-cooling has a lower transfer efficiency the procedure could be performed when the magnet needs to be filled immediately and the slow pre-cool hasn't been done because none of the Liquifiers are near full.

...