Versions Compared

Key

  • This line was added.
  • This line was removed.
  • Formatting was changed.

...

The fast pre-cooling transfer efficiencies are about 47-49% which does not allow filling the entire magnet volume using just one Liquifier. Nevertheless, higher efficiencies of about 54% can be obtained with the fast pre-cool if the Liquifier pressure is reduced to 3.5 PSI at the expense of nearly 2 hours (~8L) where the Liquifier is not producing while reaching the set pressure. Although the fast pre-cooling has a lower transfer efficiency the procedure could be performed when the magnet needs to be filled immediately and the slow pre-cool hasn't been done because any none of the Liquifiers are near full.

...