There are three instances where you can hear this warning

    • Case 1 : "Lost EIP signal!!!" on the VI. But EIP on the camera shows frequency numbers

     Action needed: Stop the VI by pressing "Stop VI". Restart the VI by pressing the "Run" button

    • Case 2: "Lost EIP signal!!!" on the VI.  EIP on the camera is not showing frequency numbers. There is no fault warning on the microwave power supply in the control room rack.

...

              • Enter 1 degree in the VI and press the right arrow to move the motor by 1 degree. Keep pressing until you see a signal. If it is not showing anything after 5 presses, change the value to 5 degrees and repeat.  If it is not successful, enter 10 degrees and try.  

    • Case 3: "Lost EIP signal!!!" on the VI.  EIP on the camera is not showing frequency numbers. The microwave power supply in the control room rack is in a faulty condition. 

...