Action needed: Slightly change the motor position until you see the EIP signal

              • Enter 1 degree in the VI and press the right arrow to move the motor by 1 degree. Keep pressing until you see a signal. If it is not showing anything after 5 presses, change the value to 5 degrees and repeat.  If it is not successful, enter 10 degrees and try.  


    • Case 3: "Lost EIP signal!!!" on the VI.  EIP on the camera is not showing frequency numbers. The microwave power supply in the control room rack is in a faulty condition. 

Action needed:

              1. Reset the power supply fault state 
                • Press the left button to switch to standby mode. Press it again to Transmit. This will reset the fault.  

If not, try following 

                • Press the left button to switch to standby mode. Press the right button two times (The ode will change from Local → Remort→ Local). Press the left button to switch to Transmit

If not, try following 

                • Powercycle the microwave power supply in the slow controls rack using the red toggle switch and Change the mode to Remort. Change the mode to Remort.  Press the left button to switch to Transmit


2. Now you will see frequency numbers on the EIP. If not slightly change the motor position until you see the EIP signal (As in the Case 2 action)