On the strip charts,
1) Ensure both Liquefiers are producing LHe. Check the top-right plot in the https://e906-gat1.fnal.gov/data-summary/e1039/target-par-preset/target-tv.php.
2) Ensure the purifier has LN2 level above 75%. Check the top-middle plot in the https://e906-gat1.fnal.gov/data-summary/e1039/target-par-preset/target-tv.php.
    If the level is below 75%, then perform a refill using the procedure: docdb 10337
3) Check the difference between the parameters "Outside gHe tank(psig)" and "PT501 (A or B or both)" using https://e906-gat1.fnal.gov/data-summary/e1039/target-all-auto.php
   
    If the difference is above 65 psig, then that indicates that the purifier needs cleaning / regeneration.
    Follow Cleaning QT purifier Helium-space (pumping-purging the purifier) procedure with the Target Expert.
    



4. Roots pumps setup

5. Checking two gHe bottles’ levels on the west-wall of the cryoplatform



6. Outside gaseous helium tank

Image

If this pressure reading gets below 45 psig, then IMMEDIATELY check with the Target Expert. You can check this pressure remotely using the strip charts.
https://e906-gat1.fnal.gov/data-summary/e1039/target-all-auto.php

If there is a need to arrange the gHe tube trailer to refill, please follow the steps on How to order LN2 refill to the tank, gHe tube-trailer, LN gHe cylinders, and LN2 portable Dewars at FNAL


7. Outside Liquid Nitrogen (LN2) tank

Image

If this level is below 24 In.H2O, then please follow the steps on How to order LN2 refill to the tank, gHe tube-trailer, LN gHe cylinders, and LN2 portable Dewars at FNAL to arrange a refill.

8. Gaseous helium, nitrogen cylinders supplies and Liquid Nitrogen portable dewars supply
Image ModifiedImage Modified

There should be at least two cylinders from each gas type (gHe and gN2), and two portable LN2 dewars available at all times.
Please follow the steps mentioned in How to order LN2 refill to the tank, gHe tube-trailer, LN gHe cylinders, and LN2 portable Dewars at FNAL to proceed with ordering.


9. Fill Procedures

1.  Filling LN2 to the purifier using the outside LN2 tank
2. Filling Magnet Shield with LN2