  • The insulation vacuum of the magnet has been pumped down for multiple days and the vacuum level is below 8x10−6 Torr
  • The U-Tube connection between the magnet and the fridge is installed and pumped down to below 1x10−3 Torr
  • The QT transfer line and all 3 stingers, Dewars and cold boxes have been pumped down to below 1x10−3 Torr
  • QT transfer line stinger is installed on the north riser of the magnet
  • Both QT dewars are full
  • Magnet is backfilled and magnet pressure is above the atmosphere
  • Magnet shield is filled with LN2
  • Purity of the system (after roots, outside tank, inlet of HR3, magnet return) is close (within 0.03%) to to reference (100.43%)


Training: No special training is needed.

...

Risks: No Identified Safety Risks

Initial valve positions

VJVA, VJVB, VJVC

Closed

VPC

Open

VJVT

Closed

Fridge run and bypass

Closed

MFC2 and MFC1

Closed

RP4

OFF

Magnet return bypass valve

0%

Action needed before 1 hour

...