ValveState
Vliq1Open
V19Closed
Vliq2Closed
V18This is a 3-way hand valve: position→ gas flow downwards
VEVACClose
V30Open
V35Open
VPClose
V36Close
REGPOpen (fully)
V11Open
Vs4Close
VfillClose
VmupClose
V17Close
V14Open
NV2Close
SVPDAlways open (relay was taken out)
V12Close
V13Close
VDeprClose
FI01Completely Opened
FI02Open (somewhere in the middle)
FI03Completely Closed
MV101-HeOpen (from tank)
MV1020-HeOpen (to tank)
MV103-HeClose


The MV1 VM1 valve which is a red box in the back of HR3 will start blinking and close the main recovery path and then it forces helium gas through REG4 (making the horn sound).  Closing this valve increases the pressure in the recovery manifold so that PC1 with a spike (usually just to 2 psi or so but less than 4 psi).  Building pressure in this space pushes helium to the buffer tank through REG4 and REG3.

...

When we actually started trying to liquefy HR3 started running normally on its own.  There may or may not be a correlation.


Information about some 'key' (or major) valves on the top part of HR3


VM1 has two operational configurations:
  1. Manual: Adjust the hand valve in the bottom to open/close
  2. Auto: Automatic control with the inputs from PC1 and PSW4.
ImageImage Added

REG4: According to the P&ID this valve regulates a flow

ImageImage Added



Kaeser Compressor


Kaeser Compressor Display (Click this link to see images)
Kaeser Compressor Maintenance

...