ValveFunctionality
REGPRegulate the gHe pressure (indicated by P10) from the tanks outside
P10Pressure indicator: shows the regulated pressure from the outside gHe tank
V30Controls gHe inlet to the purifier (There is a solenoid valve after this hand valve "V9a" should be opened from QT-HMI screen if you want to let gHe flow-in to the purifier)
V35Controls gHe outlet from the purifier (There is a solenoid valve after this hand valve "V21a" should be opened from QT-HMI screen if you want to let gHe flow-out from the purifier)
P41

Pressure transducer (connected to a manual gauge at the HR3 front panel): shows the inlet pressure to the purifier (should be equal to P10 if V30 is open)

P40Pressure transducer (connected to a manual gauge at the HR3 front panel): shows the outlet pressure from the purifier (if |P41 - P40| > 40 psig, then it is an indication that the purifier needs regeneration)
V36An isolation valve between the inlet-side and outlet-side of the purifier
V17AA bleeder valve to bleed gHe from the outlet (return) side of the gHe tank which includes the spaces : Buffer tank, filters F1, F2, F3 upstream to the compressor's outlet
V14An isolation valve for the buffer tank from the outside tank
SV7A solenoid valve is controlled by the small HMI screen on the upper-front section of the HR3.
Operational settings: Closing at 50 psig, Opening at 90 psig.







HR3 control box: normal operation configuration (06/16/2023)