1. QT portion of the HR3

Updated-HR3.pdf

ValveFunctionality
REGPRegulate the gHe pressure (indicated by P10) from the tanks outside
P10Pressure indicator: shows the regulated pressure from the outside gHe tank
V30Controls gHe inlet to the purifier (There is a solenoid valve after this hand valve "V9a" should be opened from QT-HMI screen if you want to let gHe flow-in to the purifier)
V35Controls gHe outlet from the purifier (There is a solenoid valve after this hand valve "V21a" should be opened from QT-HMI screen if you want to let gHe flow-out from the purifier)
P41

Pressure transducer (connected to a manual gauge at the HR3 front panel): shows the inlet pressure to the purifier (should be equal to P10 if V30 is open)

P40Pressure transducer (connected to a manual gauge at the HR3 front panel): shows the outlet pressure from the purifier (if |P41 - P40| > 40 psig, then it is an indication that the purifier needs regeneration)
V36An isolation valve between the inlet-side and outlet-side of the purifier
V17AA bleeder valve to bleed gHe from the outlet (return) side of the gHe tank which includes the spaces : Buffer tank, filters F1, F2, F3 upstream to the compressor's outlet
V14An isolation valve for the buffer tank from the outside tank









HR3 control box: normal operation configuration (06/16/2023)