  1. Close MV101 (supply) and MV102 (return) 
  2. Connect a He cylinder to L3 (Helium purge inlet) - Regulate 10 psig

Image AddedImage Added

      3. Open Vmup

      4. Ensure V30, V35, V9a, V21a, V11A, V14, SV03, REGX are open

Image Added Image Added

      5. Open VHSV and the rotatable valve on the top of the line

      6. Check the gauge and make sure that it reads 10 psig

      7. Start HR3

Image Added

      8. Slowly open V17 and vent


Measuring the purity:

Measure from SVSAMPLE or at the vent near V17

Image Added



Turning OFF HR3 after purging:

  1. Press the Emergency Stop and right after that close V17 (Flow on FI01 will be zero)

Image Added

     2. Close SVSAMPLE

     3. Close VHSV and the rotatable valve on top of the line

     4. Close Vmup

     5. Close the bottle