Types of Alarms

  • Microwave Control
    • It generates visible/audible alarm.
    • Expected responses are described in the operation manual.
  • Lifter Control
    • It generates visible/audible alarm.
    • Expected responses are described in the operation manual.
  • Cryo Control Panel
    • It generates visible alarm.
    • Expected responses are described below.
  • Active Alarm for Target Operator
    • The dedicated webpage should be opened on "uva-nuc-02":  https://e906-gat1.fnal.gov/data-summary/e1039/target-par-preset/target_operator_alarm.php
      • "E1039 Data Summary" → "Presets" under "Target Control Data" → "Target Operator" under "Active Alarm"
    • It generates visible/audible alarm.
    • You can confirm that this page is functioning, by enabling the "Force alarm on" checkbox () and clicking the "Update" button ().
    • Expected responses are described below.

Responses

Responses to all alarms should be listed here, or included in alarm message.

SystemAlarm Type/MessageResponse
Microwave ControlAllPlease refer to the operation manual.
Lifter ControlAllPlease refer to the operation manual.

Cryo Control Panel (CCP)

A Sub Panel status is "Black".The sub panel is not running.  Follow this instruction to start it.


A Sub Panel status is "Red".The sub panel has an error.  Follow this instruction to restart it.

Active Alarm

"Cryo Pressure_IVC" is NG.Call the target expert.  The Insulation Vacuum Chamber (IVC) might be degraded.


"Cryo Pressure_PTMR3" or "...Magnet Dewar Pressure" is NG.Call the target expert.  The target magnet might be quenched.


"Acnet_F:ANM4LCWT1" is NG.Call the target expert.


"DLLiqA0_Liquefier A PV_TI5**" or "DLLiqB0_Liquefier B PV_TI5**" is NG.Call the target expert.  Or call directly MCR to turn off KMag (NM4AN)??


"Cryo LCW_QT*" is NG.Call the target expert.