Steps:
Note: DO NOT CONNECT the purity meter to the white tube until you reach to Step (6).
- Connect the regulator (with the relief valve) at the tube trailer, and perform a leak-check with snoop.
- Open the main hand valve from the tubes, regulate ~50psig, and open the green valve of that portable manifold connected at the tube trailer.
- Check whether the rotatable hand valve is closed (see the picture).
- Open MV11 and then bleed from MV12 a few times (open-close MV12 multiple times to flush), then leave the MV12 cracked open (to have only a little flow).
- Connect the white tube's one end to the port after MV12 and let gHe flush through the other end before connecting to the purity meter.
- Connect the other end of the white tube to the purity meter inlet while there is a little gHe flow.
- Measure the purity, and take a picture of the reading, send it to the target-expert (or to the target-group), make a elog entry (with date, time, location, and purity).