Operations

Turning On

  1. Switch the main AC power to "ON", using the black handle.
  2. You might find that WH7000-1,2,3 are red due to "VFD FAILURE".  In this case, touch the "PROCESS" button and the "E-STOP" button (which should be "RESET" in red).  WH7000-1,2,3 should become green and the "E-STOP" button becomes "OK".
     
  3. Touch the "MANUAL CONTROL" button.  The password can be found on the left top corner of the cabinet.

Turning Off

a

Specifications

SV630 BF Coolant Flow Rate: 12.2 L/m → 3.2 Gal/m

Maximum flow rate (730 L/h → 12.2 L/m) is used for worst case senario. More info at SV630 BF Manual



WH 7000 (Upper Pump) Coolant Flow Rate: 5.7 L/m → 1.51 Gal/m. More info at WH 7000 Manual

Maintenance

Roots pumps oil types: PDF
Rotary vane pump oil type (recommended by Leybold Tech): LVO 130

Parts list

  1. https://www.solbergmfg.com/Solberg/media/Solberg/Combined%20Datasheets/STS-combined-series-US0621.pdf
  2. https://www.doigcorp.com/NUMATICS-F901G-16Z