Remote Magnet Controls

 

As of May 2nd, the magnet power supply will be controlled via an RDP desktop. There is a shortcut to "magnet" on the remmina icon at the upper right of the screen. Here's a quick list of the commands of this interface as it differs from PDP.

 

As of May 10th, we are leaving the magnet in driven mode, meaning the switch heater is always on and the leads are always connected to the magnet!!

 

 

 

 

 

 

 

Ramp Up Procedure

 

Restrictions:

 

 

Procedure: Instructions from Counting House which assumes: The magnet power supply is on and a sweep for magnetic objects from an area within 4m of the magnet has already been done

 

  1. Put PDP in "Monitor," if necessary, to allow updates from users
  2. Press "Unlock Magnet Controls" button in the Polarization Display Panel, PDP
  3. Press "Hold" button
  4. Hit "Heater On" button and confirm this action in the dialog box that presents itself
  5. The magnet must be ramped up in steps with the correct current rates, PDP should not allow you to exceed these rates, but be mindful. For 2.5T Set the Setpoint to 60.9135A and Setrate to 3.0A/min. There is only one step in this case. For 5.0T there are four steps (seen below). At each step enter the Setpoint and corresponding Setrate then press To SETPOINT

 

when the current has been reached press HOLD and change value to the Setpoint and Setrate for the next step and press SETPOINT. After each change of Setpoint and Setrate check the Magnet Control display in PDP to make sure Labview is reading in correctly

 

  1. Press "Hold"
  2. Wait for a few seconds
  3. Press the "Heater Off" button
  4. To ramp down leads, press "To Zero" button
  5. Press "Lock Magnet Controls" button

 

Ramp Down Procedure

 

  1. Press "Unlock Magnet Controls" button in the Polarization Display Panel, PDP
  2. If necessary, Ramp the Power Supply to the Magnet Current
  3. Press "Heater On" button and confirm this action in the dialog box that presents itself
  4. Set the first Setpoint and Setrate values. PDP should not allow you to exceed these rates, but be mindful. With 2.5T you can just press TO ZERO. With 5.0T the magnet must be ramped down in steps. At each step enter the Setpoint and corresponding Setrate then press To SETPOINT when the current has been reached press HOLD and change value to the Setpoint and Setrate for the next step and press SETPOINT. After each change of Setpoint and Setrate check the Magnet Control display in PDP to make sure Labview is reading in correctly
  5. Press the "Hold" button
  6. Press the "Heater Off" button
  7. Press "Lock Magnet Controls" button