Operations
The following procedures suppose that you use the remote control interface (i.e. Cryo Control Panel; CCP). When you use the front panel of the magnet PS, some button operations are different.
Things to-do
- We need a power-indicator on the shim power supply. These are the shim-coil-power indicators.
- We need a field indicator in the cave.
Precautions
No operation can be made when the magnet is being filled with LHe.- The magnet LHe level is so high (>20%??) that LHe remains even if soft quench happens.
- A time to ramp up/down the magnet should be planned in advance, since it takes at least 1.5 hours.
- You should remember that any deviation from the standard procedure might cause quench.
- You are expected to know what should happen at each step and confirm that it happened. You had better go through this procedure with another person who has already carried it out.
- You should search for any indication of quench during the operations, such as error messages on the magnet PS, sudden changes of Tank B, Coil T, Coil B, the magnet pressure, etc.
- You can use this chart preset: https://e906-gat1.fnal.gov/data-summary/e1039/target-par-preset/magnet_ramp.php
(A) To Ramp Up
(B) To Switch to Persistent Mode
(C) To Switch from Persistent Mode
(D) To Ramp Down
- Check whether the magnet is in the persistent mode. If so first switch back from the persistent mode by following the previous section.
- Set "0.5 A/m" and "63 A".
- Press "To Set" and wait until the current becomes the SP and "HOLD". It takes 24 minutes.
→ after ~24 minutes →
If you are in the NM4 Hall, and you see that the display shows "Rate Limit" in RED, then wait for a few minutes and observe the strip charts and this module's display.
This indicates that the Mercury-iPS is more cautious with the rate of 'current' and it is adjusted by itself.
After a few minutes, the display should look like the following.
Wait until the VI indicates that it switched to "HOLD"
- Set "1 A/m" and "45 A".
- Press "To Set" and wait until the current becomes the SP and "HOLD". It takes 18 minutes.
→ ~ 18 minutes → - Set "2 A/m" and "0 A", then Press "To Set". Then, the screen should be similar to the following screen.
- Wait until the current becomes the SP and "HOLD". It takes 23 minutes. Then, the screen should be similar to the following screen.
- Press the Toggle button of the main switch heater. It turns off both the main and shim switch heaters. The following screenshot was taken during the process after pressing the toggle button.
At the end, if you can access the NM4 hall, then ensure that the two LED bulbs Z1 and ZY-Z2 are OFF. Also ensure the Mercury-IPS shows ZEROs for Current(A), Magnet(A), Voltage(V) as shown in the following picture.
Reference: UVA elog 174
Documents
- ORC 2049 - ORC of Super conducting magnet electronic connectors (in the target cave)
- oxford_seppo_magnet_cryostat_handbook.pdf
- oxford_seppo_work_notes.pdf
t.b.w.