- Shut Down and Restart Procedures
- Helium Liquefier (QT) System Summary
- How to fill LN2 to the Purifier and Magnet LN2 spaces
- How to fix the QT-HMI if its screen got frozen
- Fixing dewar level indicator
- How to restart the QT PLC
- How to fix if the outlet flows of Liquefiers are showing strange values
- HR3 operations
- Filling LN2 to the purifier using a portable Dewar
- QT System Starting Procedure
- Liquefier Cleaning when Dewar contains LHe
- Cleaning QT purifier Helium-space (pumping-purging the purifier). Cleaning QT purifier Helium-space (OLD)
- Cleaning QT purifier LN2 space
- Fixing dewar level indicator (related to (9))
- Dewar Cleaning Procedure (using the transfer line)
- Cold-helium Circulation Procedure (Cooldown Transferline)
- Cleaning gHe outside tank using the purifier
- Resolving Liquefier Blocking Alarm when Liquefier is on Liquefy Mode
- Resolving Liquefier Blocking Alarm when Liquefier is on Idle Mode
- Purifier LN2 Fill Procedure-updated-labels.pdf ( Please use this updated one, or the original one on docdb 10337 by Jordan Bohn)
...
To check LCW parameters: https://e906-gat1.fnal.gov/data-summary/e1039/slow-cont-acnet.php
(click on NM4LCW or select )
QT LCW requires are 7.8kW of cooling per coldhead compressor, will less than 77F on water inlet and at least 6L/min of flow.
To check HMI parameters: https://e906-gat1.fnal.gov/data-summary/e1039/target-QT-HMI.php (select:
...